Hach sc1000 LXV402.99.01002

Print
The Hach sc1000 Multi-parameter Universal Controller is a fully modular system consisting of a Display Module and one or more Probe Modules.

Probe Module-Each sc1000 Probe Module provides power to the system and can accept up to 8 digital sensors. Probe Modules can be networked together to accommodate many more sensors attached to the same network.
Description

Features / Benefits

Modular System
The Hach sc1000 Multi-parameter Universal Controller is a fully modular system consisting of a Display Module and one or more Probe Modules. 

Probe Module-Each sc1000 Probe Module provides power to the system and can accept up to 8 digital sensors. Probe Modules can be networked together to accommodate many more sensors attached to the same network.

Display Module-The sc1000 Display Module is a full-featured controller with a large color touch-screen display. The intuitive, easy to use interface can be used for any number of parameters. One Display Module controls either a single Probe Module or a number of Probe Modules connected by a digital network. The Display Module is fully portable and can be disconnected and moved anywhere within the network.

"Plug and Play" Operation 
There´s no complicated wiring or set up procedures with the sc1000 controller. Just plug the sensor into a Probe Module and it´s ready for use without special ordering or software configuration.

Mix and Match Digital Sensors
The digital sensors that can be plugged into the Probe Module can be any from Hach´s line of digital sensors- dissolved oxygen, pH, ORP, conductivity, turbidity, suspended solids, nitrate, etc. Use them in any combination. 

Flexible Communication Options 
Communication and relay options for the sc1000 controller can be configured to suit any situation. Standard configurations for a single Probe Module include;

  • up to 4 potential free relay contacts for alarm and control functions,
  • up to 12 analog outputs for measured values,
  • up to 12 digital or analog inputs from instruments (i.e., flow or pressure sensors),
  • field bus card to integrate with an external network (MODBUS® and PROFIBUS DP are currently available).

Additional relays and analog inputs and outputs can be added by networking a second Probe Module or optional DIN-rail communication modules.

Digital Reliability and Integration 
Digital signals between the sc1000 controller and attached sensors assure data integrity and immunity from signal interference. Digital outputs from the sc1000 make it easy to integrate the controller into an existing network. Additional advanced communication features include:
·... Ethernet port (standard)-attach a computer directly to the sc1000 controller to operate the system from the computer. Download data logs and upload software updates.
·... MultiMediaCard (MMC) slot (standard)-transfer data and software updates using a tandard memory card like those used in digital cameras.
·... GSM wireless modem (optional)-use it for fully remote operation of the sc1000 controller, including transfer of data and software updates. (FCC approval pending.)

Expandable and Upgradeable
The sc1000 controller can adapt to your needs. Change probes without changing the controller. Probe Modules can be added or removed depending on operational needs. Fully upgradeable software ensures that this system will not be obsolete.

Specifications

Ambient Conditions
Operation: -20 to 55°C (-4 to 131°F);
0 to 95% relative humidity, non-condensing

Storage: -20 to 70°C (-4 to 158°F); 
0 to 95% relative humidity, non-condensing

Power Requirements
100 to 230 Vac, 50/60 Hz

Power: 75 W
Optional: 24 Vdc

Display
1/4 VGA graphical backlit TFT color touch screen 
Resolution: 320 x 240 pixels

Relays
Up to four SPDT, user-configurable contacts rated 100 to 230 Vac, 5 Amp resistive maximum, per probe module. Additional relays are available via digital network connection.

Outputs
Up to 12 analog 0/4-20 mA, maximum impedance 500 Ohms per probe module. 
Additional analog outputs are available via digital network connection. 
Optional digital communications via MODBUS® (RS-485) or PROFIBUS DP.

Inputs
Up to 12 analog 0-20 mA, maximum impedance 500 Ohms per probe module.
Additional inputs are available via digital network connection.

Control
PID, high/low phasing, setpoint, deadband, overfeed timer, off delay, and on delay

Alarms
Low alarm point, low alarm point deadband, high alarm point, high alarm point deadband, off delay, and on delay

Communication (Optional)
MODBUS® (RS-485): Advanced communications/networking with PLC or SCADA system directly from analyzer.

PROFIBUS DP

GSM cellular module (FCC approved)

Ethernet service port (standard)

Multimedia memory card slot (standard)

Memory Backup
All user settings are retained indefinitely in memory (non-volatile) (EEPROM)

Mounting Configurations
Surface, panel, and pipe (horizontal and vertical)

Enclosure
IP65; ABS (display module) and metal (probe module) enclosure with corrosion-resistant finish

Dimensions
Probe module with attached display module: 315 x 250 x 142 mm (12.4 x 9.8 x 5.6 in.)

Weight
Approximately 6.5 kg (14.3 lbs.) depending on configuration

Certifications
cTUVus to UL 61010A-1 and CSA C22.2 No. 1010.1
TUV-GS to EN 61010-1
CE per 73/23/EEC and 89/336/EEC

*Specifications subject to change without notice.